Engineering, 05.05.2020 06:52 fluffy374747

A gas leak has led to the presence of 1.00 mole% carbon monoxide in a 350-m^3 laboratory. The leak was discovered and sealed, and the laboratory is to be purged with clean air to a point at which the air contains less than the OSHA (Occupational Safety and Health Administration)-specified Permissible Exposure Level (PEL) of 35 ppm (molar basis). Assume that the clean air and the air in the laboratory are at the same temperature and pressure and that the laboratory air is perfectly mixed throughout the purging process.

a. Let tr (h) be the time required for the specified reduction in the carbon monoxide concentration. Write a differential CO mole balance, letting N equal the total moles of gas in the room (assume constant), x the mole fraction of CO in the room air, and v (m/h) the flow rate of purge air entering the room (and also the flow rate of laboratory air leaving the room). Convert the balance into an equation for dxldt and provide an initial condition.
b. Integrate the balance to derive an equation for tr in terms of vp.
c. If the volumetric flow rate is 700 m^3/h (representing a turnover of two room volumes per hour), how long will the purge take? What would the volumetric flow rate have to be to cut the purge time in half?

ansver
Answers: 2

Another question on Engineering

question
Engineering, 04.07.2019 18:10
Coiled springs ought to be very strong and stiff. si3n4 is a strong, stiff material. would you select this material for a spring? explain.
Answers: 2
question
Engineering, 06.07.2019 03:10
Two slings are used to lift a girder of length l and uniform weight per unit length. what is the best position for the slings if the bending moment due to the weight of the girder is to be kept to a minimum?
Answers: 2
question
Engineering, 06.07.2019 03:20
Asimply supported beam abc of length l has ab of length hl and second moment of area / and bc of length hl and second moment of area i2. determine the strain energy stored in the beam when it is subject to a vertical load f at its midpoint and the consequential central deflection.
Answers: 2
question
Engineering, 06.07.2019 03:30
The catalogue for a refrigerant 22, four-cylinder, hermetic compressor operating at 29 r/s. a condensing temperature of 40°c and an evaporating temperature of -4°c shows a refrigeration capacity of 115 kw. at this operating point, the motor (its efficiency is 90 percent) draws 34.5 kw. the bore of the cylinders is 87 mm and the piston stroke is 70 mm. the performance data are based on 8°c of sub-cooling of the liquid leaving the condenser. compute (a) the actual volumetric efficiency and (b) the compression efficiency.
Answers: 2
You know the right answer?
A gas leak has led to the presence of 1.00 mole% carbon monoxide in a 350-m^3 laboratory. The leak w...
Questions
question
Social Studies, 02.02.2021 21:30
question
Mathematics, 02.02.2021 21:30
question
Mathematics, 02.02.2021 21:30
question
Physics, 02.02.2021 21:30
question
Mathematics, 02.02.2021 21:30
question
Advanced Placement (AP), 02.02.2021 21:30
Questions on the website: 13722367